5 – Installation and set-up
MultiExact Gas Analyser
5.5.4 Automatic validation/calibration target and tolerance
1.
Select
➞
Set up
➞
Auto val.
2.
Use the
and
soft keys
to highlight the measurement for
which you want to set up the
validation/calibration, then press the
soft key.
Figure 29 – The auto val/cal screen
3.
Select target, the auto val/cal low
target value screen will then be
displayed.
Figure 30 – The auto val/cal
parameters screen
4.
Edit the displayed low target value as
described in Section 4.10.
Figure 31 – The auto val/cal
low target value screen
☞
The low target value can only be changed for the O
2
trace, purity and control
measurements (see Section 2.5).
5.
Edit the low tolerance value.
Figure 32 – The auto val/cal
low tolerance screen
46
05410001A / Revision 6