MultiExact Gas Analyser
5 – Installation and set-up
Flushing time
Specifies the flushing time (0 min 30 sec to 59 min
59 sec)
Timer
On (active) or off (inactive). If active the calibration or
validation can be carried out regularly at a specified
frequency.
Start time/date
Only applicable if the timer is active. This specifies the
time/date on which the first automatic validation/
calibration will be carried out. See Section 5.4.3.
Repeat time
Only applicable if the timer is active. If set (in the range
1 hour to 999 days), specifies the interval between
successive automatic calibrations/validations. Select
0 hours and 0 days if repeat is not required.
Figure 28 shows a typical automatic calibration sequence, where:
•
Pre-warning is on.
•
Automatic calibration of transducer 1 (Tx1) has been initiated.
•
validation/calibration type has been set to low & high.
•
Mode has been set to calibration.
•
Inerting is set to on (active) and inert gas has been set to low.
05410001A / Revision 6