MultiExact Gas Analyser
6 – Calibration
7.
If low calibration was successful the
screen should display as shown in
Figure 86.
8.
Adjust the valve so that the
measurement reading is between
105 and 115% O
2
.
9.
Adjust your calibration gas supply so
that the flow rate is the same rate as
noted in Step 6.
Figure 86 – Pressure high
calibration screen
10.
Allow the O
2
measurement shown on the screen to stabilize for at least
30 seconds, then press the
soft key.
11.
Open the valve or remove completely and adjust the calibration gas supply so
that the flow is the same as noted in step 6. Pressure transducer calibration is
now complete.
6.4
Calibrating the flow alarm (option)
If the flow alarm option has been fitted you must calibrate the flow alarm when you first
calibrate the analyser and annually thereafter.
Before commencing the calibration procedure, ensure the ambient temperature has
been relatively constant for at least half an hour.
Use the following procedure to calibrate the flow alarm:
1.
Select
➞
Service
➞
Flow alarm.
2.
Select the zero calibration option and
stop the flow of gas through the
analyser. Wait 3 minutes before
pressing the
soft key.
Figure 87 – The flow calibrate
zero screen
3.
Select the normal calibration option
and restore the sample flow through
the analyser. Wait 3 minutes before
pressing the
soft key.
Figure 88 – The flow calibrate
normal screen