MultiExact Gas Analyser
5 – Installation and set-up
5.4.5 Adjusting the backlight timer
When the analyser is first switched on, the backlight goes on to illuminate the screen.
If no soft key is pressed, the backlight will remain on for the preset 'backlight time', and
will then go off. The timer associated with the backlight time is restarted whenever a
soft key is pressed (that is, the backlight remains on for the backlight time after the last
soft key press). To adjust the backlight time:
1.
Select
➞
Settings
➞
Backlight.
2.
Change the backlight time (duration)
setting as required, then press the
soft key.
Figure 23 – The backlight
timer screen
☞
The backlight time (duration) can be set between 0 and 999 seconds. Set
the backlight time (duration) to 0 seconds to leave the backlight permanently
switched on.
5.4.6 Setting the clock
☞
The time and date will remain set for approximately 1 week if the power supply
to the analyser is switched off.
Use the following procedure to set the date and time:
1.
Select
➞
Settings
➞
Clock.
2.
Edit the displayed time as described
in Section 4.10.
Figure 24 – The clock (time) screen
3.
Edit the displayed date as described
in Section 4.10.
4.
The date format can be set to your
regional preference ('day/month/
year' or 'month/day/year'):
refer to Section 5.4.7.
Figure 25 – The clock (date) screen