Model PIM-3 13-5
#01W5150
<ret>
#01W6pounds
<ret>
#01W72.997
<ret>
#01W8111.768
<ret>
#01W91
<ret>
Third, the analog to digital converter inside the PIM-3
is internally calibrated to minimize temperature errors.
This may be done regardless of the load on the trans-
ducer.
#01F3
<ret>
Next, the tare value will be entered into the PIM-3. All
load must be removed from the transducer before this
command is performed so that the PIM-3 will recognize
this point as "zero".
#01F1
<ret>
Then, calibrate the system using the shunt cal value.
Make certain that there is still no load being applied to the
transducer.
#01F4
<ret>
This completes the system shunt calibration.