Model PIM-3 11-7
11.6
F5
Apply Shunt Resistor and Transmit Read-
ing
SYNOPSIS
F5
- Applies the Shunt Resistor and transmits the reading
DESCRIPTION
The
F5
function applies the shunt resistor (usually 59,000 ohms for SENSO-
TEC transducers) on the amplifier channel to two transducer leads (usually -
SIG and -SENSE for SENSOTEC transducers) and transmits the reading.
EXAMPLE
#00F5
The above command, when followed by a carriage return, will cause the
instrument addressed as "00" to apply the shunt resistor to the transducer
and transmit the reading.
WHEN TO USE IT
This can be used as quick check to verify that a strain gauge transducer is
connected and operating properly.