4-6 Mounting and Wiring
4.3
Shunt Cal Resistor
If a SENSOTEC transducer has been purchased and
shipped with the PIM-3, the proper shunt calibration resis-
tor has been installed. If not, SENSOTEC normally ships
a 59,000 ohm 0.01% resistor installed in PIM-3. If the
transducer used with PIM-3 requires another value of
resistor, it may be necessary to remove the 59,000 ohm
resistor and install another. To install another shunt cal
resistor, proceed as follows:
1. Remove mating connectors from PIM-3, and remove the cover
of the unit. The printed circuit boards will remain attached to the lid.
2. Locate the shunt calibration resistor. It is located on the ana-
log circuit board (furthest from the lid), near the transducer connector
location. The resistor is mounted in two pin jacks for easy removal.
Using a pair of long-nose pliers, gently remove the resistor, and
replace it with the proper value for your transducer.
3. Carefully replace the lid on PIM-3 and screw it in place.
Содержание PIM-3
Страница 1: ...2080 Arlingate Lane Columbus Ohio 43228 614 850 5000 DIGITAL INLINE AMPLIFIER MODEL PIM 3 ...
Страница 6: ...NOTES ...
Страница 10: ...1 4 Introduction ...
Страница 16: ...3 4 Block Diagram ...
Страница 18: ...4 2 Mounting and Wiring ...
Страница 23: ...Model PIM 3 4 7 ...
Страница 24: ...4 8 Mounting and Wiring ...
Страница 32: ...6 6 Command Format ...
Страница 35: ...Model PIM 3 7 3 ...
Страница 36: ...7 4 RS 232 Installation Notes ...
Страница 46: ...9 4 Application Programs Figure 9 1 Windows 3 1 Terminal Communications Settings ...
Страница 58: ...10 12 System Commands ...
Страница 74: ...11 16 Amplifier Commands ...
Страница 77: ...Model PIM 3 12 3 ...
Страница 79: ...Model PIM 3 12 5 ...
Страница 97: ...A 8 Appendix ...