Model PIM-3 11-15
11.14
WK
Write Known Load Calibration Value
RK
Read Known Load Calibration Value
SYNOPSIS
WK
{value} - Change Known load calibration value to {value}
DESCRIPTION
The
WK
command changes the known load calibration value. The numeric
value given in the
information field
is in the engineering units for that ampli-
fier channel.
When the PIM-3 is calibrated to the transducer using the "Known Load” this
is the value that should be applied to the transducer during calibration.
The PIM-3 will acknowledge the write command by responding with "
OK
".
The
RK
command is used to read the known load calibration value.
EXAMPLE
The command below, when followed by a carriage return, will change the
known load calibration value to 20000.
#00WK20000
WHEN TO USE IT
The known load calibration value must be changed to match the precision
load that you have available to calibrate your transducers. This value is only
used when calibrating with the
F8
command.
Содержание PIM-3
Страница 1: ...2080 Arlingate Lane Columbus Ohio 43228 614 850 5000 DIGITAL INLINE AMPLIFIER MODEL PIM 3 ...
Страница 6: ...NOTES ...
Страница 10: ...1 4 Introduction ...
Страница 16: ...3 4 Block Diagram ...
Страница 18: ...4 2 Mounting and Wiring ...
Страница 23: ...Model PIM 3 4 7 ...
Страница 24: ...4 8 Mounting and Wiring ...
Страница 32: ...6 6 Command Format ...
Страница 35: ...Model PIM 3 7 3 ...
Страница 36: ...7 4 RS 232 Installation Notes ...
Страница 46: ...9 4 Application Programs Figure 9 1 Windows 3 1 Terminal Communications Settings ...
Страница 58: ...10 12 System Commands ...
Страница 74: ...11 16 Amplifier Commands ...
Страница 77: ...Model PIM 3 12 3 ...
Страница 79: ...Model PIM 3 12 5 ...
Страница 97: ...A 8 Appendix ...