Model PIM-3 11-3
11.2
F1
Activate Tare/Zero
SYNOPSIS
F1
- Activates Tare/Zero operation
DESCRIPTION
The Tare function forces the presently applied load on the transducer to be
read as zero.
To deactivate the Tare/Zero function, you can send the
F2
command.
The PIM-3 will not send a response after this command has been processed.
This command is a universal command. Sending this command for the spe-
cial address of “FF” will cause all PIM-3’s on the communications link to pro-
cess the command.
EXAMPLE
#00F1
The above command, when followed by a carriage return, will cause the
instrument addressed as "00" to activate the Tare/Zero function and force
the data to read zero.