TROUBLESHOOTING
Problem
Checks / Corrective Action
Measurements not
taking place
- Check reported on-board temperature is within limits
- Check supply voltage is within limits (reported in main log)
- Check laser temperature within limits (reported in Oryx
system log)
- Check all ‘Power’ LEDs are illuminated during a
measurement
- Check Main log and Oryx system log for other errors
Temperature trace
has excessive noise
or does not read
correctly to the end
of the fibre
- Clean connectors
- Re-splice fibre joints
- Check cable run for sharp bends / breaks
- Check fibre types use in cable run
- Point losses and a fiber break can easily be identified from
the raw signal display. Typical connector loss = 0.3dB
Typical fusion splice loss = 0.1dB (one way)
Temperature trace
exhibits step
changes at fibre
joints
- Clean connectors
- Re-splice fibre joints
- Use double-ended method
Unstable calibration
– temperature trace
jumps excessively
after each
measurement
Incorrect slope or
offset on
temperature trace
- Check reference sections have been chosen
appropriately. Avoid reference sections across a
temperature step
- Try using fixed calibration constants
- Clean connectors
Error saving data /
writing to file or disk
- Ensure there are no spaces at the end of Installation or
Channel names
- De-select read only permissions for all files and folders in
C:\Sensornet DTS and all files and folders in the Oryx
ROOT folder
- ODC should be run under administrator privileges
No communications
with Oryx
- Check PC serial com port settings (Oryx default:
115200b/s; 8bit; 1 stop bit; no flow control
- Check Oryx COM port settings
Содержание ORYX DTS
Страница 1: ...This page intentionally blank ...
Страница 2: ......
Страница 5: ......
Страница 6: ...This page intentionally blank ...
Страница 70: ...Page 64 Oryx DTS User Manual v4 Figure 45 Setting a zone Figure 46 The list of defined zones ...
Страница 100: ...Page 94 Oryx DTS User Manual v4 This page intentionally blank ...
Страница 101: ......