Oryx DTS User Manual v4
Page 9
Figure 4 - Connecting and disconnecting the E2000 optical connector
NOTE: Ensure the optical connector is pushed firmly into the housing and latches to
obtain proper mating
Typical connector loss is 0.1 – 0.3dB. Values above 0.4dB would warrant cleaning
the connectors. The fibre optic connector panel can be removed by unscrewing the
four knurled thumb screws and carefully removing the panel. There is sufficient
slack on the inside fibre pigtails to allow removal and cleaning.
CAUTION! If one of the internal optical connectors is damaged,
the unit will need repair at Sensornet. Only remove the Optical
Connector Panel if a dirty connector has been inserted or the
optical connector loss cannot be reduced by cleaning the
external connector.
In cases where the DTS is connected and disconnected frequently (e.g. if the Oryx
is transported to different sites for measurements), the use of an external ‘sacrificial’