Page 52
Oryx DTS User Manual v4
•
Y zoom – zoom on Y axis only
•
Full scale – zoom out to show full scale on both axes
•
Zoom out – zoom out from cursor position
•
Zoom in – zoom in from cursor position
It is also possible to simply double-click the start and end scale legend values on
the graphical plot and enter numerical value to more precisely set the scales.
ORYX CONFIGURATION WIZARD
Collecting data
The Configuration Wizard is used to create a new configuration file whenever the
Oryx DTS system is connected to a new installation. Configuration files contain
important calibration information for an individual installations ensuring the DTS can
accurately measure temperature. The DTS system is pre-calibrated for use with
common types of 50/125 graded index multimode optical fibre. As such, it is not
essential to run the Configuration Wizard but is recommended for fine tuning the
system for a specific installation.
To begin the Configuration Wizard, select “Measurement Config Wizard” from the
ODC main menu (Figure 16). The user is prompted to select the particular Oryx
required and then presented with the first wizard screen to allow an existing
configuration to be loaded if required. Select “Load File…” if an existing
configuration is to be loaded and modified, otherwise click “Next”. The user is then
prompted to enter a name for this configuration (Installation Name) and optionally,
any notes associated with this (Figure 35).
Содержание ORYX DTS
Страница 1: ...This page intentionally blank ...
Страница 2: ......
Страница 5: ......
Страница 6: ...This page intentionally blank ...
Страница 70: ...Page 64 Oryx DTS User Manual v4 Figure 45 Setting a zone Figure 46 The list of defined zones ...
Страница 100: ...Page 94 Oryx DTS User Manual v4 This page intentionally blank ...
Страница 101: ......