STP-H803/H1303 Series Instruction Manual
◇
Tuning can be performed only while the STP pump is in the
LEVITATION state (It cannot be performed while the STP pump is in
the ACCELERATION, NORMAL OPERATION, or BRAKE state).
◇
When performing the tuning during remote operation, use the
"RESET" switch on the front panel after switching the "REMOTE"
switch to the manual. (The remote signal cannot be used for tuning.)
◇
Once tuning is completed, re-tuning is not required unless the
configuration (the STP pump serial number, the STP control unit
serial number, and the STP connection cable length) is changed.
O
pe
ra
tio
n
◇
Although the STP pump sounds during tuning, this is not an
indication of abnormality.
9.1.2
Each of data after tuning
The specific information and the user setting are recorded in the STP pump and
STP control unit.
When tuning after changing the configuration (the STP pump serial number, the
STP control unit serial number), the STP pump data overrides the STP control
unit data except control unit serial No. and total hours of running (control unit).
Table 9.1 shows the data recorded in the STP control unit.
The user setting may be changed when replacing the STP pump. Confirm the
setting after tuning, and reconfigure the setting when the user setting has
changed.
Table 9.1 Data List
Data Data
priority
Reconfiguration
Pump serial No.
Pump
Control unit serial No.
Control unit
Total hours of running (Pump)
Pump
Total hours of running (Control unit)
Control unit
Number of starts
Pump
Error record
Pump
Unnecessary
(specific
information)
Remote operation mode/Detailed
Pump
Rated speed
Pump
Valid/invalid TMS function
Pump
Valid/invalid of inhibit function
Pump
Valid/invalid of Emergency vent.
Pump
Necessary
(user setting)
9-2