STP-H803/H1303 Series Instruction Manual
11.6.1
Cleaning and Decontamination
The method of cleaning and decontamination of the STP pump and STP control
unit is shown below.
Table 11.5 Cleaning and Decontamination
Area
Cleaning and Decontamination
Exterior
Wipe with proper solvent (such as alcohol).
Clean off with a pure gas when dust has accumulated in the
connector.
When the label of the STP pump has been damaged, contact
Service office.
STP Pump
Interior
The decontamination (overhaul) in the STP pump is executed
in Service office. When overhaul is needed, contact Service
office.
Exterior
Wipe with proper solvent (such as alcohol).
When dust has accumulated in the ventilation port, wipe off or
vacuum it with the cleaner. In this case, dust must not enter
in the STP control unit.
Clean off with a pure gas when dust has accumulated in the
connector.
When the label of the STP control unit has been damaged,
contact Service office.
STP
Control unit
Interior
The cleaning (overhaul) in the STP control unit is executed in
Service office. When overhaul is needed, contact Service
office.