STP-H803/H1303 Series Instruction Manual
11.4.2
Second Damage Limit
"Second Damage Limit" is displayed when continuing to operate the STP pump
after "First Damage Limit" is displayed, and the accumulation of the damage
point attains the second setting value by generating second "Disturbance" or
"Mains Failure". Overhaul the STP pump because the STP pump cannot
perform the rotational operation when "Second Damage Limit" is displayed.
11.5
"ERROR RECORD" Message Display Function
The STP control unit can display the contents of up to the last 10 errors on the
LCD.
To view the error record, push the "SELECT" switch follow the instructions for
entering Confirmation Mode in Section 9. The error number of up to the last 10
errors is displayed on the upper line of the LCD in the form "Error Record n
°
*/*",
and the contents of the error is displayed on the lower line.
The first number of "**/**" shows the number of the error (the most recent error
is number 1) and the second number shows the number of errors that have
occurred up to 10. Thus, a "1/3" would denote the most recent error of 3 errors.
Press the "UP" or "DOWN" switch to scroll display the contents of the next or
previous error in the error record. (The "UP" and "DOWN" switch can also be
input in Remote Mode. See Section 9.3 for a more detailed explanation of
Confirmation Mode.
Troubleshooting