Delay Time
:
0 –999, minimum resolution: 100 μs
Calibration Function
:
Open / Short calibration with full frequency or dot frequency, Load
Parameters Displayed
:
Value, DABS, D%
Parameter display of test page :
One set main/ sub parameter, 10 point list seep
List Sweep Function
:
10 points list sweep Frequency, AC voltage/current, internal/external bias
voltage/ current can be swept. Each sweep point can be sorted
separately
Comparator Function
:
10 bin sorting, BIN1– BIN9, NG, AUX, Bin counter PASS/FAIL on front
panel, LED indication
Calibration
Warm Up Time
:
> 30 mins.
Ambient Temperature
:
23 + 5o C
Test Signal Voltage
:
0.3 Vrms – 1 Vrms
Zeroing
:
After Open or Short
General Information
Display
:
4.3” TFT LCD display
Internal Non-volatile Memory
:
100 sets of LCRZ setting files, 201 times test results
External USB memory
:
GIF files LCRZ setting files Test data can be stored via USB memory
directly.
Interface
:
USB, HANDLER, RS232C, LAN, USB Host, GPIB (optional),
DCI (optional)
Operating conditions
:
0 °C - 40 °C, £ 90% RH
Supply
:
230 V + 10 % , 50 Hz AC
Power Consumption
:
Max. 80 VA
Dimensions (WxHxD)
:
325 x 90 x 370 mm ,
370 x 108 x 410 mm (with handle & sheath)
Weight
:
Approx. 5.7 kg
(subject to change without notice)