User Manual SM6026
69
<BIN COUNT MEAS>
The current page is the bin count display page.
<LIST SWEEP MEAS>
The current page is the list sweep display page.
<MEAS SETUP>
The current page is the measurement setup page.
<CORRECTION>
The current page is the correction page.
<LIMIT TABLE SETUP>
The current page is the limit table setup page.
<LIST SWEEP SETUP>
The current page is the list sweep setup page.
<SYSTEM SETUP>
The current page is the system setup page.
<FILE LIST>
The current page is the file list page.
The :LINE command is used to set the current measurement item which can be a substring with up to 16
characters. The :LINE? query returns the current measurement item. The character string of the
measurement item can be used as the file name when saving a file.
Command syntax: DISPlay:LINE”<string>”
Where,
<string> can be an ASCII character string (maximum number is 16).
For example: WrtCmd(“MEAS:LINE” Resistor meas””)
Query syntax: DISPlay:LINE?
Return format: <string><NL^END>
The :ResultFONt command is used to set the current font of the measurement result. The :ResultFONt?
Query returns the current font of the measurement result.
Command syntax: DISPlay:RFONt <font>
<font> can be the following information:
LARGe: Use large character to display the measurement result, 12ms/meas.
TINY: Use tiny character to display the measurement result, 5ms/meas.
OFF: Measurement result will not be displayed but can be read from the bus.
Query syntax: MDISPlay:FRONt?
Return format: <font><NL^END>
<font> can be the following information:
LARGE
TINY
OFF
7.1.2
FREQuency subsystem commands
The FREQuency subsystem commands are mainly used to set the measurement frequency of the
instrument. The :FREQuency query returns the current measurement frequency.
<value>
Command syntax:
FREQuency
MIN
MAX
Where,
<value>
NR1, NR2 or NR3 data format followed by Hz, kHz, MHz.
MIN
Set the measurement frequency as 20Hz.
MAX
Set the measurement frequency as 200kHz (The maximum frequency of SM6026 A is
300kHz)
For example: WrtCmd(“FREQ 1KHZ”)
Set the frequency as 1000Hz.
Query syntax: FREQuency?
Return format: <NR3><NL^END>
Содержание SM6026
Страница 1: ...User Manual LCR Meter SM6026 ...