The following control parameters can be set on the
<BIN COUNT>
page.
Count function ON/OFF (COUNT)
There are 2 zones are displayed in this page: BIN COUNT DISP, COUNT. The function of each zone will
be introduced as below.
The following test result/ condition can be displayed in this page but cannot be set in this page. Users can
set them on the
<LIMIT TABLE SETUP>
page.
Test parameter (PARAM)
Nominal value (NOM)
Bin limit value (HIGH/ LOW)
3.3.1
PARAM
Parameter zone shows the “Function” parameter, if user selects primary and secondary parameter swap
compare mode, the parameter will be displayed as swap parameter, such as: “Cp-D” is displayed as
“D-Cp”, which means the current D is compared as primary parameter while Cp is compared as
secondary Parameter.
3.3.2NOM
.
Nominal parameter is the nominal value used to make bin compare.
3.3.3BIN
This zone shows the bin number of the limit list. “2nd” means the secondary parameter limit.
3.3.4HIGH/LOW
This zone shows the high and the low limits of the limit list.
3.3.5COUNT
This zone shows the count value of the current bin.
3.3.6AUX
This zone shows the count value of the auxiliary bin.
3.3.7OUT
This zone shows the count value of the out bin.
Operation steps for bin count function
Execute the following operations to set the bin count function ON/OFF on
<BIN COUNT DISP>
page.
1) On < BIN COUNT DISP> page, move the cursor to COUNT zone, the following soft keys will be
displayed.
ON
OFF
RESET
2) Press the soft key ON to turn on the count function.
3) Press the soft key OFF to turn off the count function.
4) Press the soft key RESET, “☺ : Reset count, Sure?” will be displayed in the help zone. Then the
following soft keys will be displayed.
YES
NO
5) Press the soft key YES to reset all bin counts to 0.
6) Press the soft key NO to cancel the reset operation.