User Manual SM6026
70
7.1.3
VOLTage subsystem commands
The VOLTage subsystem commands are mainly used to set the measurement voltage. The VOLTage?
query returns the current measurement voltage.
Command syntax:
<value>
VOLTage
MIN
MAX
Where,
<value>
NR1, NR2 or NR3 data format followed by V.
MIN
Set the measurement voltage as 5mV.
MAX
Set the measurement voltage as 2V.
For example: WrtCmd(“VOLT 1V”) Set the measurement voltage as 1V.
Query syntax: VOLTage?
Return format: <NR3><NL^END>
7.1.4
CURRent subsystem commands
The CURRent subsystem commands are mainly used to set the measurement current. The CURRent?
query returns the current measurement current.
Command syntax:
<value>
CURRent
MIN
MAX
Where,
<value>
NR1, NR2 or NR3 data format followed by MA.
MIN
Set the measurement current as 50μA.
MAX
Set the measurement current as 20mA.
For example: WrtCmd (“CURR 10MA”) Set the measurement current as 10mA.
Query syntax: CURRent?
Return format:<NR3><NL^END>
7.1.5
AMPLitude subsystem commands
The :AMPLitude subsystem commands are mainly used to set the auto level control (ALC) function as
ON or OFF. The :AMPLiture? query returns the current status of the ALC function.
Command systax:
ON
AMPLitude:
ALC OFF
1
0
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd (“AMPL:ALC 0”) Set the ALC function as OFF.
Query syntax: AMPLitude:ALC?