User Manual SM6026
39
If the primary parameter is out of the limit range, it is sorted into BIN OUT. If the primary parameter of
DUT is within the limit range but its secondary parameter is out of the limit range, the DUT will be sorted
into the Aux bin.
——————————————————————————————————————————————
Note: When the secondary parameter only has low limit and the auxiliary bin is set as ON, if the primary
parameter of DUT is within the limit range and the secondary parameter is smaller than or equal to its low
limit, the DUT will be sorted into the auxiliary bin. When the secondary parameter only has high limit and
the auxiliary bin is set as ON, if the primary parameter of DUT is within the limit range and the secondary
parameter is larger than or equal to its high limit, the DUT will be sorted into the auxiliary bin.
——————————————————————————————————————————————
Operation steps for setting the auxiliary bin function ON/OFF
:
1) Move the cursor to AUX, the following soft keys will be displayed.
ON
OFF
2) Use above soft keys to set the auxiliary function as ON or OFF.
3.7.6
HIGH/LOW
SM6026
can set bin limits of 9 primary parameters and one secondary parameter.
The test results
can be sorted into 10 bins at most (BIN 1 to BIN 9 and BIN OUT). The high/low limits of primary
parameters can be set in high limit and low limit of bins from BIN 1 to BIN 9. The limit of the secondary
parameter can be set in
HIGH
and
LOW
of 2nd.
Operation steps for setting high/low limit
:
Execute the following steps to set sorting limits.
1) Set
PARAM
and
NOM
in the compare function menu and the limit MODE of the primary parameter.
2) Move the cursor to Low limit of BIN 1. If you select tolerance mode, the following operation steps
should be from step 3 to step 6; if you select sequential mode, the following operation steps should be
from step 7 to step 11.
3) User numeric keys to input low limit value in Low limit. After inputting the data, you can use (
p, n, μ, m,
k, M, *1
) to replace
[ENTER]
to input the limit value. When [ENTER] is used, the default unit is the
same as that being input last time. When pressing
*1
, the default unit will be F, H or Ω. After inputting
limit value in
LOW
of BIN 1, the low limit of BIN 1 will be automatically set as – (absolute limit) and the
high limit will be + (absolute limit).
4) The cursor will automatically move to
LOW
of BIN 2. Repeat step 3 until the limits of BIN 9 are input.
Then the cursor will automatically move
LOW
of 2
nd
.
5) After inputting the low limit of the secondary parameter, the cursor will automatically move to
HIGH
of
2
nd
.
6) Input the high limit of the secondary parameter.
7) In Low limit of BIN 1, use numeric keys to input the low limit. After inputting the data, you can use (
p, n,
μ, m, k, M, *1
) to replace
[ENTER]
to input the limit value. When
[ENTER]
is used to input the limit
value, the default unit is the same as that being input last time. When pressing
*1
, the default unit will
be F, H or Ω.
8) After inputting the low limit of BIN 1, the cursor will automatically move to
HIGH of BIN 1
. Input the
high limit of
BIN 1
.
9) The cursor will automatically move
HIGH
of BIN 2. For the limit mode is sequential mode, the low limit
of BIN 2 will be the high limit of BIN 1. Input the high limit of BIN 2.