User Manual SM6026
29
2) Use above soft keys to set the average times or use numeric keys and
[ENTER]
input average times
directly.
3.5.5
Level monitor function
The level monitor function can monitor the real voltage across DUT or real current through DUT. The
monitored voltage value is displayed in
Vm
zone on
<MEASURE DISP>
page while the monitored
current value is in
Im
zone.
——————————————————————————————————————————————
Note: The correction function can influence the level monitor function, so when the correction data
changes the level monitor value will change. When the correction is switched between OPEN or SHOR
or LOAD, the level monitor value will be influenced as well
——————————————————————————————————————————————
Operation steps for setting the level monitor function
:
Execute the following operation steps to set the level monitor function as ON or OFF.
1) Move the cursor to Vm zone, the following soft keys will be displayed.
ON
OFF
2) Press ON to set the voltage monitor function as ON while press OFF to set the voltage monitor
function as OFF.
3) Move the cursor to Im zone, the following soft keys will be displayed.
ON
OFF
4) Press ON to set the current level monitor function as ON while press OFF to set the current level
monitor function as OFF.
3.5.6
Delay time
SM6026
trigger delay means the delay time from triggering to test-start. Delay function can set the
trigger delay time. When the list sweep test function is used, all set delay time will be delayed at each
sweep test point. The range of the trigger delay time can be set from 0s to 60s with 1ms as the resolution.
The trigger delay function is great useful when the instrument is applied in an auto test system. When the
instrument is triggered by HANDLER interface, the trigger delay time can ensure DUT and test terminal
has a reliable contact.
Operation steps for setting the delay function
:
Execute the following steps to set the measurement delay time.
1) Move the cursor to the
DELAY
zone.
2) Use numeric keys to input delay time. After pressing a numeric key, the following unit keys will be
displayed. These soft keys can replace
[ENTER]
to input delay time.
msec
sec
3.5.7
Output impedance
SM6026
provides four output impedance for your choice: 100 Ω, 50 Ω, 30 Ω and 10Ω. When testing
inductance, it is necessary to input the same output impedance so as to make data comparison with
other instruments.