  Power Connection 

§

  Check the voltage rating and the plug 

design. 

$

  Make sure that the voltage rating printed 

on the manufacturer’s ID label is identi-
cal to that of your local line voltage.  
If the voltage specified on the label or 
the plug design of the AC adapter do not 
match your country’s rating or standard, 
please contact the nearest Sartorius 
office or your equipment dealer. 

 

The power connection must be made in 
accordance with the regulations appli-
cable in your country. The device (pro-
tection class 1) must be plugged into a 
properly installed wall outlet which has 
a protective grounding conductor (PE) 

  Connecting the Analyzer to AC Power

 

Note:

 

The heating unit of the moisture ana-
lyzer has been factory-set to 230 V or 
120V, depending on the information 
given in your order. The voltage setting 
is indicated on the manufacturer’s label; 
for example:

–  230 volts: LMA200PM-...EU 
–  115 volts: LMA200PM-...US

!

If the voltage setting is not correct,  
do not operate the moisture analyzer! 
The equipment must be modified by  
a Sartorius service technician.

–  Use only original power cords approved 

by Sartorius or by a trained electri-

cian. If the cable supplied with the 
AC  adapter is not long enough, make 
sure to use an extension cord that has 
a grounding conductor. If you use an 
electrical outlet that does not have a 
protective grounding conductor, make 
sure to have an equivalent protective 
conductor installed by a certified elec-
trician as specified in the applicable 
 regulations for installation in your 
country. Make sure the protective 
grounding effect is not neutralized by 
use of an extension cord that lacks a 
protective grounding conductor.

and a fuse with maximum 16A. The 
power plug or other suitable device for 
disconnecting the power must be easily 
accessible so the LMA200PM can be 
disconnected quickly in case of hazard.

  Safety Precautions

 

If you use an electrical outlet that  
does not have a protective grounding 
conductor, make sure to have an equiv-
alent protective conductor installed by  
a certified electrician as specified in the 
applicable regulations for installation in 
your country. Make sure the protective 
grounding effect is not neutralized by 
use of an extension cord that lacks  
a protective grounding conductor.

6

  Switching on the LMA200PM

§

  Press the on/off switch (17) to switch on the moisture analyzer.

  Equipment not in Use

 

If the equipment will remain unused for only a short period, press the 

 key (5) to switch 

it to standby mode. 
 
If it will remain unused for a long period, use the on/off switch (17) to switch it off.

Warmup Time

To deliver exact results, the weighing system in the LMA200PM  
moisture analyzer must warm up for at least 30 minutes after initial 
connection to AC power or after a relatively long power outage.  
Only after this time will the system have reached the required operating 
temperature.

  Connecting  

the Power Cord 

§

  Plug the power cord into 

a wall outlet (mains) 
that has been installed 
in accordance with 
industry standards and 
is equipped with  
a protective ground 
(earth) conductor (PE).

  Connecting Electronic Peripheral 

Devices 

§

  Make absolutely sure to unplug the 

scale from AC power before you con-
nect or disconnect a peripheral device 
(printer, scanner or PC) to or from the 
interface port. 

Содержание LMA200PM

Страница 1: ...Operating Instructions Sartorius Microwave Moisture Analyzer Model LMA200PM 98648 015 60 98648 015 60...

Страница 2: ...disinfect Symbols The following symbols are used in these instructions Instructions for proper operation Instructions for exceptional cases The outcome of an operating step Warning hazard Intended Use...

Страница 3: ...re the cooling fans are always operational In the event of defect switch off the moisture analyzer and inform service personnel Warnings and Safety Precautions If you use cables purchased from another...

Страница 4: ...mber Do not use a sample container that prevents the cover from closing completely The moisture analyzer must not be operated if the sample cham ber is not completely closed and sealed by the analyzer...

Страница 5: ...secondary raw material Equipment Supplied LMA200PM microwave moisture analyzer model LMA200PM 000EU or LMA200PM 000US Adapter with sample retainer Power cord Glass fiber pads 80 Printer paper 1 roll P...

Страница 6: ...e the protective grounding effect is not neutralized by use of an extension cord that lacks a protective grounding conductor and a fuse with maximum 16A The power plug or other suitable device for dis...

Страница 7: ...ve function Navigating the menu Returns to previous menu level During input Cancels input Confirm During measurement Activates the selected function Navigating the menu Shows next menu level During in...

Страница 8: ...Measured value result line This line shows the weight readout with the weight unit at stability or a calculated value a plus or minus sign and the weight unit displayed Text lines The text lines show...

Страница 9: ...Tap the second touchscreen button from the left the arrow pointing towards the left 9 Entering data using the touchscreen keypad If no is shown the value can be entered or edited using the touchscree...

Страница 10: ...w of the Moisture Analyzer on page 4 Built in printer Ethernet interface RS 232 interface Built in Printer You can configure printer settings and printout formats including settings for ISO GLP confor...

Страница 11: ...ap the Login button First login Admin Administrator Subsequent login list of all users Tap the First login Administrator line Admin PIN Subsequent login Name PIN Tap the name PIN line Enter your PIN F...

Страница 12: ...this example Jones Group Operator and press the key PIN 4 Select the Supervisor group Tap the button Group Supervisor and press the key 5 Open the PIN input window Tap the button and press the key 6 E...

Страница 13: ...Results are displayed as follows Each dataset is shown on two lines First line Internal number program name date time Second line User name sample ID result with unit and analysis time There are a num...

Страница 14: ...ic printing Headers 1 to 3 Print formats You can set the following parameters for printing data records and statistics Data record header Data record intermediate results Data record footer Statistics...

Страница 15: ...erature Info Shows information about the LMA200PM the weighing module and the databases To print an informational screen Press the Print key 15 Overview of User Privileges Group Administrator Supervis...

Страница 16: ...page 17 Results Sample ID See page 19 Evaluation of results Printing parameters Automatic printout Header 1 Header 2 Header 3 Print formats Header Intermediate results Footer Statistics See page 19 C...

Страница 17: ...nfo only Device name Alphanum Use DHCP Option IP address nnn nnn nnn nnn Subnet mask nnn nnn nnn nnn Standard gateway nnn nnn nnn nnn Remote access for VNC client nnn nnn nnn nnn Operating parameters...

Страница 18: ...formats Setup Configuration Device Serial Built in On interfaces printer Off RS 232 Off Printout Protocol XON XOFF parameters RTS CTS 2 wire No protocol Baud rate 300 bd 600 bd 1 200 bd 2 400 bd 4 80...

Страница 19: ...termediate Blank line Header 1 results Header 2 Date and time Header 3 Time GLP header ID1 Date and time ID Time Analysis time Result Operator ID1 Footer Blank line ID2 Program name Date time Drying p...

Страница 20: ...0 1111 to 9999 repeat PIN input User privileges SET Calibration Supervisor SET Manage programs SET Eval of results SET Print param SET Printout formats SET Date and time SET Serial ports SET Ethernet...

Страница 21: ...range Analysis mode Analysis mode Standard Addition of water Concentration mg l Drying process Standard Level 2 to 100 2 stages Level 1 2 to 100 30 to 100 Time 1 0 0 to 10 0 min Level 2 2 to 100 30 to...

Страница 22: ...01 0 00 Result display mode Moisture m Conversion factor 1 0000 to 2 0000 Solids S Conversion offset 50 000 to Volatile V 50 000 ppm moist ppmM ppm solids ppmS Ratio MS Weight loss mg Residue g Create...

Страница 23: ...nto stages with different microwave output in each stage Note The time of the last stage is the deter mined by the shutoff parameter in all drying processes Standard The microwave output remains const...

Страница 24: ...oist ppmM ppm solids ppmS Ratio MS Weight loss mg Residue g You can define the following for the Moisture M and Dry weight S display modes Conversion factor Conversion offset If one of these parameter...

Страница 25: ...th information is shown briefly i Loading The display s info line shows LMA200PM Rel 01 00 00 Prg Ketchup 5 Begin moisture analysis Tap the Analyze button 28 11 2007 14 41 38 0 000 g Prg Ketchup Mode...

Страница 26: ...11 2007 16 41 46 Sartorius Microwave Moisture Analyzer LMA200PM 000EU Rel 01 00 00 Serial no 12345678 Prg Ketchup Mode Standard Process R Standard Level 30 End Automatic Initial weight 2 125 g During...

Страница 27: ...on the weighing module Remove any pad or sample from the sample retainer 5 Tare the weighing module Tap the Tare button Status line yellow background Taring the weighing module 6 Calibrate the weighin...

Страница 28: ...rrent program or Create new program the bar graph shows abso lute values 0 to 70 g Checkweighing graph If Minimum maximum or Setpt tolerance is selected for the Initial weight parameter the bar graph...

Страница 29: ...ions Menu lines are not touch sensitive Navigation and selection are possible only using the touchscreen buttons at the bottom of the display Press the key to insert selection Press the key to return...

Страница 30: ...orius cables with order number 7357312 or 7357315 RS 232 Interface Type of interface Serial interface Operating mode Full duplex Standard RS 232 Protocol XON XOFF RTS CTS 2 wire or none Transmission r...

Страница 31: ...egardless of the labels in the above example the arrow keys shown on the touchscreen buttons at any given point The following equivalents also apply F5 Back F12 Print Enter Confirm Arrow keys Like the...

Страница 32: ...IP address of computer 192 168 178 20 IP address of LMA200PM 192 168 178 21 The settings for subnet mask and standard gateway should be the same as those in the local computer These settings are confi...

Страница 33: ...ot be Allow the scale to warm up completed within the required time period again and repeat the adjustment expected xxx Wrong format Enter data correctly Wrong format bad xxx illegal xxx Outside the r...

Страница 34: ...fter analysis Open the cover Use the forceps to grasp the glass fiber pad and carefully remove it from the sample chamber Replacing the Printer Paper Press down briefly on the cover of the printer com...

Страница 35: ...particular the sealing surfaces between the upper and lower parts of the sample chamber is not permitted as this will damage the PTFE coating If any part of the sealing surfaces between the upper and...

Страница 36: ...30 V 50 Hz 1200 VA LMA200PM 000EU 115 V 60 Hz 1200 VA LMA200PM 000US Fuse 2 10 A T 5 20 mm Ambient conditions Indoor use only Altitude up to 2000 m Operating temperature range 5 C to 40 C 41 F to 104...

Страница 37: ...37 All dimensions given in mm Dimensions Scale Drawings...

Страница 38: ...38...

Страница 39: ...orms are used to facilitate legibility in these instructions and always simultaneously denote the other gender as well Copyright notice This instruction manual including all of its components is prote...

Отзывы: