![Sartorius LMA200PM Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/sartorius/lma200pm/lma200pm_operating-instructions-manual_1199622036.webp)
36
Specifications
Measurement method
Sample heated by microwave radiation; moisture content calculated from loss on drying
Microwave generator
1000 W
Power output
2% to 100%, configurable in 1%-steps
Measurement range
8% to 100%
Accuracy
From 1 g initial sample /– 0.05%
Response time
Approximately 40 to 120 seconds (depends on sample and moisture content)
Display of results
% moisture content with calculation options; ppm moisture content,
% volatile components, %, ppm solids, g solids, mg solids, % ratio
Weighing capacity
70 g
Resolution | reading accuracy
0.1 mg
of the weighing system
Reading accuracy of the measurement results
0.001%
Display module
4
VGA
Control elements
Touchscreen, 5 function keys
Program memory
320 drying programs with alphanumeric designation
Data storage
The 500 most recent results, with statistical evaluation, stored in battery-backed memory
Built-in printer
Thermal printer, 40 characters per line
Data interfaces
1 serial RS-232 interface, 1 Ethernet port
Power source
230 V, 50 Hz, 1200 VA (LMA200PM-000EU)
115 V, 60 Hz, 1200 VA (LMA200PM-000US)
Fuse
2
+
10 A/T (5
+
20 mm)
Ambient conditions
Indoor use only
Altitude
up to 2000 m
Operating temperature range
5°C to 40°C (41°F to 104°F)
Allowable storage temperature range
–10°C to +60°C (14°F to 140°F)
Humidity
Maximum relative humidity: 80% for temperatures up to 31°C,
linear decrease to 50% relative humidity at +40°C
Dimensions (H
+
W
+
D)
304
+
510
+
535 mm (12
+
20
+
21 in.)
Weight
Approximately 22 kg (48.4 lbs)
Pollution degree 2
Normally only nonconductive pollution or temporary conductivity caused by condensation.
Limitation of emissions
In accordance with EN 61326-1 (IEC 61326-1):
Group 1, Class B, suitable for use in domestic establishments and establishments
directly connected to a low-voltage power-supply network that supplies buildings used
for domestic purposes.
Immunity to interference
In accordance with EN 61326-1, (IEC-1 ):
Immunity test requirements for equipment intended for use in industrial locations (Table 2)
Electrical safety
In accordance with EN 61010-1 (IEC 61010-1)
Accessories/Options
Product
Order No .
80 glass fiber pads
6906940
500 disposable pipettes
YAT01MA
Printer paper (five 20-m rolls)
69MA30100
Specifications