Setting the Drying Parameters
Purpose
Adapt the moisture analyzer’s software to the special requirements of products.
Parameters can be configured individually for each program.
Drying Parameters (Overview)
Setup
Input
Manage
Load programs
programs*
Change
Short name
Short name
Alphanum.
current
Additional identification
Long name
Alphanum.
program
ID 1
Alphanum.
ID 2
Alphanum.
Microwave mode
Microwave mode
High range
Low range
Analysis mode
Analysis mode
Standard
Addition of water
Concentration (mg/l)
Drying process
Standard
Level
2 to 100%
2 stages
Level 1**
2 to 100% (30 to 100%)
Time 1
0.0 to 10.0 min
Level 2**
2 to 100% (30 to 100%)
3 stages
Level 1**
2 to 100% (30 to 100%)
Time 1
0.0 to 10.0 min
Level 2**
2 to 100% (30 to 100%)
Time 2
0.0 to 10.0 min
Level 3**
2 to 100% (30 to 100%)
Initial weight
Off
Minimum/maximum
Start
Unlocked
Locked
Min
0.01 g to 69.99 g
Max.
Min. + 0.01 g to 70.00 g
Target value/tolerance
Start
Unlocked
Locked
Target value
0.01 g to 63.00 g
Tolerance
1.0 to 50%
Sample capturing
Standard
Fast
Very fast
Key without stability
21
Operation