Error messages are shown in the text line or measured value line of the display.
Display
Cause
Solution
No segments shown
No AC power is available
Check the AC power supply
AC adapter not plugged in
Connect AC adapter to the
wall outlet (mains)
Fuses are defective
Replace the fuses
The analyzer is switched off
Press the power switch (on/off)
on the back panel
L / ERR 54
Load is below the weighing range
Place a load on the weighing module
Sample retainer not in place
Place the sample retainer on the
weighing module
H
The load exceeds the weighing
Unload the weighing module
module capacity
ERR 02
Calibration/adjustment condition
Calibrate only when zero is displayed
not met, e.g.,
– not tared
Tare the weighing module
– the sample retainer is loaded
Unload the sample retainer
ERR 03
Calibration/adjustment could not be
Allow the scale to warm up
completed within the required time period
again and repeat the adjustment
expected xxx
Wrong format
Enter data correctly
Wrong format
bad xxx/illegal xxx
Outside the range
Enter data correctly
Not permitted
must not be more than xxx
Outside the range
Enter data correctly
Too high
must not be less than xxx
Outside the range
Enter data correctly
Too low
Value exceeds display
Value has more characters
Change configuration.
than the display can show
See the chapter entitled “Configuration,”
under “Configuring Parameters in the
Setup Menu”
ERR Config. data
Checksum errors; may indicate change
Switch the moisture analyzer off
in versions (e.g., after software update) or
and on again
ERR Adjust. data
dead battery
ERR Printout config.
ERR DB user
ERR DB programs
ERR DB results
Messages
Messages are shown on the display at the time the event triggering the message occurs (covers other displayed data). For details on
messages, see the chapter entitled “Operating Design.”
Troubleshooting Guide