  Features

  Load programs*

 

The program database in the 
LMA200PM can store up to 320 drying 
programs. To view a list of available 
programs, sorted alphabetically by short 
name, select Setup > Manage programs 
> Load program.

  Change current program*

 

Short name, Long name

 

You can designate both a short name 
(up to 15 characters) and a long name 
(up to 20 characters) for each program.

 

Each short name must be unique in the 
database. 

  ID1, ID2

 

You can assign 2 identifiers (up to 20 
characters each) for each program. 

  Microwave mode

 

There are two modes to choose from for 
operation of the microwave:

–  High range (2% - 100%)
–   Low range (30% - 100%), roughly  

10% - 30% of the maximum microwave 
output. We recommend using the low 
range mode as a rule.

  Analysis mode

 

There are three modes to choose from 
for analysis:

–  Standard
–  Addition of water
–   Concentrate

 

Standard

 

Most analyses are performed using this 
method.

 

Addition of water

 

When the sample is heated by microwave 
radiation, dipolar molecules, such as 
water molecules, in the sample absorb 
the radiated energy.

 

Relatively dry samples – with a moisture 
content below 10% – contain few dipo-
lar molecules. Thus it takes longer to 
heat the sample. Adding distilled water 
to the sample has a catalytic effect,  
in that the energy from the microwave 
radiation is absorbed faster, causing the 
sample to heat up more quickly.

 

Note:

 

The added distilled water evaporates 
completely during the drying process 
and does not distort the results of  
moisture analysis.

 

Concentration

 

The LMA200PM can analyze only small 
volume samples. If the proportion of 
solids for a larger volume is sought, the 
“Concentrate" function is used.

 

Example: 

 

A 5-ml sample is taken from 1 liter of 
waste water. Before starting the analysis, 
the user enters “1000 ml" as the  
initial volume. The moisture analyz-
er multiplies the proportion of solids 
detected by a factor of 200 (1000 ml:  
5 ml). The result is the proportion of sol-
ids in 1000 ml of the sample material.

 

Drying process

 

The drying process can be divided into 
stages, with different microwave output 
in each stage.

 

 

 

Note:

 

The time of the last stage is the deter-
mined by the shutoff parameter in all 
drying processes. 

–  Standard:
 

The microwave output remains constant 
throughout the entire drying process.

–  2 stages:
 

The drying process is divided into two 
stages, with different levels of microwave 
output in each stage. You can configure 
a period of 0.0 to 10.0 minutes for first 
stage.

–  3 stages:
 

The drying process is divided into three 
stages, with different levels of microwave 
output in each stage. You can configure 
periods of 0.0 to 10.0 minutes for the 
first and second stages. 

*  Access rights required

  Initial weight

 

You can define conditions for the initial 
weight that must be met before the 
analysis can start:

–  Off:
 

No conditions are defined for the initial 
weight

–  Minimum/maximum:
 

The initial weight must be greater than 
the defined minimum weight and less 
than the defined maximum weight. 

 

If “Locked" is set under “Start," analysis  
cannot be performed if the initial 
weight is not between the minimum 
and maximum values. If “Unlocked" is 
set, the minimum and maximum values 
are merely guidelines.

–  Setpoint/tolerance limits:
 

You can define a target weight  
(setpoint) and associated tolerance  
limits for the initial weight. 

 

In this case, the initial weight must be 
greater than the setpoint minus toler-
ance and less than the setpoint plus  
tolerance. If “Locked" is set under 
“Start," analysis cannot be performed  
if the initial weight is not within the 
tolerance limits. If “Unlocked" is set, the  
setpoint and tolerance values are merely 
guidelines. 

Sample capturing

The speed at which the balance registers 
weights and transfers data during initial 
weighing can be adapted to the particu-
lar sample being analyzed:

– 

Standard

– 

Fast

– 

Very fast

– 

Key w/o stability

Standard:
For normal samples and normal  
handling;

Fast:
For samples that must be applied quick-
ly, or if you require a relatively exact 
weight toward a target;

Very fast:
For samples containing slightly volatile 
components;

Key w/o stability:
The balance performs as under the “very 
fast” option;
For samples with volatile constituents, 
such as hairspray:
Spray sample on pan, wait 1–2 sec. 
(spray pressure)

Содержание LMA200PM

Страница 1: ...Operating Instructions Sartorius Microwave Moisture Analyzer Model LMA200PM 98648 015 60 98648 015 60...

Страница 2: ...disinfect Symbols The following symbols are used in these instructions Instructions for proper operation Instructions for exceptional cases The outcome of an operating step Warning hazard Intended Use...

Страница 3: ...re the cooling fans are always operational In the event of defect switch off the moisture analyzer and inform service personnel Warnings and Safety Precautions If you use cables purchased from another...

Страница 4: ...mber Do not use a sample container that prevents the cover from closing completely The moisture analyzer must not be operated if the sample cham ber is not completely closed and sealed by the analyzer...

Страница 5: ...secondary raw material Equipment Supplied LMA200PM microwave moisture analyzer model LMA200PM 000EU or LMA200PM 000US Adapter with sample retainer Power cord Glass fiber pads 80 Printer paper 1 roll P...

Страница 6: ...e the protective grounding effect is not neutralized by use of an extension cord that lacks a protective grounding conductor and a fuse with maximum 16A The power plug or other suitable device for dis...

Страница 7: ...ve function Navigating the menu Returns to previous menu level During input Cancels input Confirm During measurement Activates the selected function Navigating the menu Shows next menu level During in...

Страница 8: ...Measured value result line This line shows the weight readout with the weight unit at stability or a calculated value a plus or minus sign and the weight unit displayed Text lines The text lines show...

Страница 9: ...Tap the second touchscreen button from the left the arrow pointing towards the left 9 Entering data using the touchscreen keypad If no is shown the value can be entered or edited using the touchscree...

Страница 10: ...w of the Moisture Analyzer on page 4 Built in printer Ethernet interface RS 232 interface Built in Printer You can configure printer settings and printout formats including settings for ISO GLP confor...

Страница 11: ...ap the Login button First login Admin Administrator Subsequent login list of all users Tap the First login Administrator line Admin PIN Subsequent login Name PIN Tap the name PIN line Enter your PIN F...

Страница 12: ...this example Jones Group Operator and press the key PIN 4 Select the Supervisor group Tap the button Group Supervisor and press the key 5 Open the PIN input window Tap the button and press the key 6 E...

Страница 13: ...Results are displayed as follows Each dataset is shown on two lines First line Internal number program name date time Second line User name sample ID result with unit and analysis time There are a num...

Страница 14: ...ic printing Headers 1 to 3 Print formats You can set the following parameters for printing data records and statistics Data record header Data record intermediate results Data record footer Statistics...

Страница 15: ...erature Info Shows information about the LMA200PM the weighing module and the databases To print an informational screen Press the Print key 15 Overview of User Privileges Group Administrator Supervis...

Страница 16: ...page 17 Results Sample ID See page 19 Evaluation of results Printing parameters Automatic printout Header 1 Header 2 Header 3 Print formats Header Intermediate results Footer Statistics See page 19 C...

Страница 17: ...nfo only Device name Alphanum Use DHCP Option IP address nnn nnn nnn nnn Subnet mask nnn nnn nnn nnn Standard gateway nnn nnn nnn nnn Remote access for VNC client nnn nnn nnn nnn Operating parameters...

Страница 18: ...formats Setup Configuration Device Serial Built in On interfaces printer Off RS 232 Off Printout Protocol XON XOFF parameters RTS CTS 2 wire No protocol Baud rate 300 bd 600 bd 1 200 bd 2 400 bd 4 80...

Страница 19: ...termediate Blank line Header 1 results Header 2 Date and time Header 3 Time GLP header ID1 Date and time ID Time Analysis time Result Operator ID1 Footer Blank line ID2 Program name Date time Drying p...

Страница 20: ...0 1111 to 9999 repeat PIN input User privileges SET Calibration Supervisor SET Manage programs SET Eval of results SET Print param SET Printout formats SET Date and time SET Serial ports SET Ethernet...

Страница 21: ...range Analysis mode Analysis mode Standard Addition of water Concentration mg l Drying process Standard Level 2 to 100 2 stages Level 1 2 to 100 30 to 100 Time 1 0 0 to 10 0 min Level 2 2 to 100 30 to...

Страница 22: ...01 0 00 Result display mode Moisture m Conversion factor 1 0000 to 2 0000 Solids S Conversion offset 50 000 to Volatile V 50 000 ppm moist ppmM ppm solids ppmS Ratio MS Weight loss mg Residue g Create...

Страница 23: ...nto stages with different microwave output in each stage Note The time of the last stage is the deter mined by the shutoff parameter in all drying processes Standard The microwave output remains const...

Страница 24: ...oist ppmM ppm solids ppmS Ratio MS Weight loss mg Residue g You can define the following for the Moisture M and Dry weight S display modes Conversion factor Conversion offset If one of these parameter...

Страница 25: ...th information is shown briefly i Loading The display s info line shows LMA200PM Rel 01 00 00 Prg Ketchup 5 Begin moisture analysis Tap the Analyze button 28 11 2007 14 41 38 0 000 g Prg Ketchup Mode...

Страница 26: ...11 2007 16 41 46 Sartorius Microwave Moisture Analyzer LMA200PM 000EU Rel 01 00 00 Serial no 12345678 Prg Ketchup Mode Standard Process R Standard Level 30 End Automatic Initial weight 2 125 g During...

Страница 27: ...on the weighing module Remove any pad or sample from the sample retainer 5 Tare the weighing module Tap the Tare button Status line yellow background Taring the weighing module 6 Calibrate the weighin...

Страница 28: ...rrent program or Create new program the bar graph shows abso lute values 0 to 70 g Checkweighing graph If Minimum maximum or Setpt tolerance is selected for the Initial weight parameter the bar graph...

Страница 29: ...ions Menu lines are not touch sensitive Navigation and selection are possible only using the touchscreen buttons at the bottom of the display Press the key to insert selection Press the key to return...

Страница 30: ...orius cables with order number 7357312 or 7357315 RS 232 Interface Type of interface Serial interface Operating mode Full duplex Standard RS 232 Protocol XON XOFF RTS CTS 2 wire or none Transmission r...

Страница 31: ...egardless of the labels in the above example the arrow keys shown on the touchscreen buttons at any given point The following equivalents also apply F5 Back F12 Print Enter Confirm Arrow keys Like the...

Страница 32: ...IP address of computer 192 168 178 20 IP address of LMA200PM 192 168 178 21 The settings for subnet mask and standard gateway should be the same as those in the local computer These settings are confi...

Страница 33: ...ot be Allow the scale to warm up completed within the required time period again and repeat the adjustment expected xxx Wrong format Enter data correctly Wrong format bad xxx illegal xxx Outside the r...

Страница 34: ...fter analysis Open the cover Use the forceps to grasp the glass fiber pad and carefully remove it from the sample chamber Replacing the Printer Paper Press down briefly on the cover of the printer com...

Страница 35: ...particular the sealing surfaces between the upper and lower parts of the sample chamber is not permitted as this will damage the PTFE coating If any part of the sealing surfaces between the upper and...

Страница 36: ...30 V 50 Hz 1200 VA LMA200PM 000EU 115 V 60 Hz 1200 VA LMA200PM 000US Fuse 2 10 A T 5 20 mm Ambient conditions Indoor use only Altitude up to 2000 m Operating temperature range 5 C to 40 C 41 F to 104...

Страница 37: ...37 All dimensions given in mm Dimensions Scale Drawings...

Страница 38: ...38...

Страница 39: ...orms are used to facilitate legibility in these instructions and always simultaneously denote the other gender as well Copyright notice This instruction manual including all of its components is prote...

Отзывы: