I-
8
BF-Sirius
Operation Manual
Push the reset switch. Beep tones stop and the scan table can not be moved manually.
Step3:
The dialog shown in Figure 3-2 can be selected only after releasing the emergency stop.
Step4:
OK
The scan table returns to the origin position.
Cancel
The scan table remains at the position where it stopped.
Put PCB into the bottom correctly and resume usual image scan operation.
Step5:
3.1.2 Area Sensor
Area sensor watches safe operation of shading procedure and image scanning.
When area sensor is sensed, beep tones sound and the motor stops immediately.
The system operation is suspended.
Upper Area Sensor
Front Area Sensor
Area Sensor
FFFure
3-3
When area sensor senses and stops, the dialog shown in Figure 3-4 appears.
Remove any object sensed by area sensor.
OK
The scan table returns to the origin position.
Cancel
The scan table remains at the position where it stopped.
Area Sensor Senses
FFFure
3-4
Содержание BF-Sirius
Страница 1: ... SJ241DCM1 02E BF Sirius Manual ...
Страница 7: ...I 1 BF Sirius Operation Manual Operation Manual ...
Страница 37: ...II 1 BF Sirius Maintenance Manual Maintenance Manual ...
Страница 61: ...BF Sirius Installation Manual III 1 Installation Manual ...