In addition to the BLOCK_ERR and XD_ERROR parameters, more detailed
information on the measurement status can be obtained via
DIAGN_DEV_ERR. Table C-10 on page C-6 lists the potential errors and the
possible corrective actions for the given values. The corrective actions are in
order of increasing system level compromises. The first step should always
be to reset the transmitter and then if the error persists, try to the steps in
Table C-10. Start with the first corrective action and then try the second.
Table C-10. Device Errors
Diagnostics
Table C-11. Temperature sensor
type
Bit Number
Value of
DIAGN_DEV_ERR
Description
Corrective Actions
0
0
No alarm active
See page 6-1
1
0x00000001
Reserved
See page 6-1
2
0x00000002
FF card to transmitter comm fault
See page 6-1
3
0x00000004
Level Measurement Failure
See page 6-1
4
0x00000008
Temperature Measurement Failure
See page 6-1
5
0x00000010
Volume Measurement Failure
See page 6-1
6
0x00000020
Invalid ATP
See page 6-1
7
0x00000040
No surface echo
See page 6-1
8
0x00000080
Tank signal clip warning
See page 6-1
9
0x00000100
Empty Tank
See page 6-1
10
0x00000200
Full Tank
See page 6-1
11
0x00000400
Conf. Reg. Pwd. Enabled
See page 6-1
12
0x00000800
DB Error
See page 6-1
13
0x00001000
Microwave unit error
See page 6-1
14
0x00002000
Display error
See page 6-1
15
0x00004000
Analog out error
See page 6-1
16
0x00008000
Other HW error
See page 6-1
17
0x00010000
Configuration error
See page 6-1
18
0x00020000
SW error
See page 6-1
19
0x00040000
DB Warning
See page 6-1
20
0x00080000
Microwave unit Warning
See page 6-1
21
0x00100000
Display Warning
See page 6-1
22
0x00200000
Analog out Warning
See page 6-1
23
0x00400000
Other HW Warning
See page 6-1
24
0x00800000
Configuration Warning
See page 6-1
25
0x01000000
SW Warning
See page 6-1
Value
TEMP_CONV_METHOD Description
0
User defined linearization table
1
User defined formula
2
DIN PT 100
3
CU90
Содержание 5600 Series
Страница 2: ......
Страница 4: ......
Страница 18: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 1 10 ...
Страница 64: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 2 46 ...
Страница 80: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 3 16 ...
Страница 99: ...Reference Manual 00809 0100 4024 Rev BA September 2005 4 19 Rosemount 5600 Series ...
Страница 100: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 4 20 ...
Страница 130: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 6 20 F OUNDATION Fieldbus ...
Страница 144: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 7 14 2210 Display Unit ...
Страница 174: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series A 20 ...
Страница 185: ...Reference Manual 00809 0100 4024 Rev BA September 2005 B 11 Rosemount 5600 Series 5600APPROVAL03 EPS ...
Страница 186: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series B 12 5600APPROVAL04 EPS ...
Страница 196: ...Reference Manual 00809 0100 4025 Rev BA April 2002 Rosemount 5600 Series C 8 F OUNDATION Fieldbus ...
Страница 204: ...Reference Manual 00809 0100 4025 Rev BA September 2005 Rosemount 5600 Series E 4 F OUNDATION Fieldbus ...
Страница 208: ...Reference Manual 00809 0100 4024 Rev BA March 2004 Rosemount 5600 Series Index 4 ...
Страница 209: ......