Reference Manual
00809-0100-4024, Rev BA
September 2005
7-3
Rosemount 5600 Series
2210 Display Unit
Operation
The Rosemount 2210 Display unit can be used for configuration as well as for
viewing tank data. The four softkeys allow you to navigate through the
different menus, and to select various functions for service and configuration.
(See Section 2: Mechanical Installation for information on how to connect the
2210 Display Unit.)
If you leave the Display Unit in Service or Setup mode without pushing any
button for 10 minutes (set in User Defined), it is automatically switched to
View mode, presenting the same measurement variable that was displayed
last time View mode was open.
The main menu contains the following options:
Figure 7-2. Main Menu
•
The
View
option allows you to view level data and signal strength.
•
The
Service
option allows you to view configuration status, edit holding
registers, reset holding registers to factory values, do a software reset
or to start a search for the surface echo.
•
The
Setup
option allows you to configure a transmitter.
•
The
Display Panel
option allows you to set units for measured values,
to set language and to change the user password.
Adjusting the LCD contrast
The LCD contrast can be increased by simultaneously pressing the two
buttons on the right-hand side. Press the two left-hand buttons to decrease
the contrast. It takes approximately 10 seconds to adjust from minimum to
maximum display panel contrast.
MAINMENU_43.EP
S
Содержание 5600 Series
Страница 2: ......
Страница 4: ......
Страница 18: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 1 10 ...
Страница 64: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 2 46 ...
Страница 80: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 3 16 ...
Страница 99: ...Reference Manual 00809 0100 4024 Rev BA September 2005 4 19 Rosemount 5600 Series ...
Страница 100: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 4 20 ...
Страница 130: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 6 20 F OUNDATION Fieldbus ...
Страница 144: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 7 14 2210 Display Unit ...
Страница 174: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series A 20 ...
Страница 185: ...Reference Manual 00809 0100 4024 Rev BA September 2005 B 11 Rosemount 5600 Series 5600APPROVAL03 EPS ...
Страница 186: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series B 12 5600APPROVAL04 EPS ...
Страница 196: ...Reference Manual 00809 0100 4025 Rev BA April 2002 Rosemount 5600 Series C 8 F OUNDATION Fieldbus ...
Страница 204: ...Reference Manual 00809 0100 4025 Rev BA September 2005 Rosemount 5600 Series E 4 F OUNDATION Fieldbus ...
Страница 208: ...Reference Manual 00809 0100 4024 Rev BA March 2004 Rosemount 5600 Series Index 4 ...
Страница 209: ......