Reference Manual
00809-0100-4024, Rev BA
September 2005
B-3
Rosemount 5600 Series
4. The product must be installed by suitably trained personnel and carried
out in accordance with all appropriate international, national, and local
standard codes of practice and site regulations for intrinsically safe
apparatus and in accordance with the instructions contained within this
manual.
5. Repair of this equipment shall be carried out by the manufacturer or in
accordance with the applicable code of practice.
6. All externally connected intrinsically safe apparatus must comply with the
specified IS entity parameters.
7. The Flameproof/Explosionproof enclosure may not be opened while
energized.
8. The certificate marking is detailed on drawing numbers 9150076-931
and 9150076-932.
9. The certificate has special conditions for safe use associated with it,
denoted by the X on the end of the certificate no., which must be
observed when the equipment is installed.
10. The certification of this equipment relies on the following materials used
in its construction:
If the equipment is likely to come into contact with aggressive substances,
then it is the responsibility of the user to take suitable precautions that
prevent it from being adversely affected, thus ensuring that the type of
protection is not compromised.
Aggressive substances - e.g. solvents that may affect polymeric materials.
Suitable precautions - e.g. regular checks as part of routine inspections or
establishing from the material’s data sheet that it is resistant to specific
chemicals.
Table B-2.
Antenna Type
Size
Tank Seal
Material
Temperature
Range
Pressure Range
(linear interpolation
between breakpoints)
Cone
Pipe
All
PTFE
-40 to +200°C
-1 to 10 bar @ -40°C
-1 to 10 bar @ 100°C
-1 to 5 bar @ 200°C
Cone
Pipe
All
Quartz
-40 to +400°C
-1 to 55 bar
Cone/purging
All
PTFE
-40 to +200°C
-1 to 10 bar
Cone/purging
All
Quartz
-40 to +400°C
-1 to 10 bar
Process Seal
4-in. / DN100
PTFE
-40 to +150°C
-1 to 5 bar @ -40°C
0 bar @ +150°C
Process Seal
6-in. / DN150
PTFE
-40 to +150°C
-1 to 2 bar @ -40°C
0 bar @ +150°C
Rod 100
Rod 250
All
PTFE
-40 to +200°C
25 bar @ -40°C
25 bar @ 100°C
16 bar @ 200°C
Parabolic
Parabolic
18-in. / Welded
18-in. / Clamped
PTFE
-40 to +230°C
-40 to +230°C
-1 to 10 bar
-0.5 to 0.5 bar
Содержание 5600 Series
Страница 2: ......
Страница 4: ......
Страница 18: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 1 10 ...
Страница 64: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 2 46 ...
Страница 80: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 3 16 ...
Страница 99: ...Reference Manual 00809 0100 4024 Rev BA September 2005 4 19 Rosemount 5600 Series ...
Страница 100: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 4 20 ...
Страница 130: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 6 20 F OUNDATION Fieldbus ...
Страница 144: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series 7 14 2210 Display Unit ...
Страница 174: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series A 20 ...
Страница 185: ...Reference Manual 00809 0100 4024 Rev BA September 2005 B 11 Rosemount 5600 Series 5600APPROVAL03 EPS ...
Страница 186: ...Reference Manual 00809 0100 4024 Rev BA September 2005 Rosemount 5600 Series B 12 5600APPROVAL04 EPS ...
Страница 196: ...Reference Manual 00809 0100 4025 Rev BA April 2002 Rosemount 5600 Series C 8 F OUNDATION Fieldbus ...
Страница 204: ...Reference Manual 00809 0100 4025 Rev BA September 2005 Rosemount 5600 Series E 4 F OUNDATION Fieldbus ...
Страница 208: ...Reference Manual 00809 0100 4024 Rev BA March 2004 Rosemount 5600 Series Index 4 ...
Страница 209: ......