R&S AM300
Arbitrary and Function Generator
Operating manual, 02/2005
6-149
E-1007.9922.12
6.2.3.1.2
Setting the Starting and Stopping Frequencies
Application
You have two possible ways of setting the sweep frequency range. In this
method you enter a starting (START) and stopping frequency (STOP).
Setting the
starting frequency
1.
In the
submenu
press the
function key
.
An entry field containing the currently applicable setting is displayed. The
default setting is "20 kHz".
2.
Enter a new value ( 5-55).
The permissible entry range for the starting frequency is dependent on
the waveform and for the
Sine
waveform, for example, is:
100 mHz
≤
Start
≤
35 MHz
in 10 µHz steps
Setting the
stopping frequency
3.
In the
submenu
press the
function key
.
An entry field containing the currently applicable setting is displayed. The
default setting is "100 kHz".
4.
Enter a new value ( 5-55).
The permissible entry range for the stopping frequency is dependent on
the waveform and for the
Sine
waveform, for example, is:
100 mHz
≤
Span
≤
35 MHz
in 10 µHz steps
The new setting is stored and then displayed in the parameter field, if
frequency sweeping is switched on ( 6-154).
Note
The upper frequency limit depends on the waveform selected ( 6-75).