R&S NRP-Z28/-Z98
Remote Control - Commands
1170.8966.12
6.33
E-4
List of Remote-Control Commands
The remote-control commands of the R&S NRP-Z28/-Z98 have a syntax based on standard SCPI
1999.0, but they comply with it only to a limited extent.
Table 6-8
List of remote-control commands
Command
Parameter
Unit
Default setting
Page
* Commands
*IDN?
*RST
*TRG
*TST?
CALibration Commands
CALibration:DATA[?]
<calibration data set as definite length
block>
CALibration:DATA:LENGth?
Bytes
CALibration:ZERO:AUTO[?]
OFF | ON | ONCE
OFF (fixed)
SENSe Commands
SENSe:AVERage:COUNt[?]
1 to 65536
4
SENSe:AVERage:COUNt:AUTO[?]
OFF | ON | ONCE
ON
SENSe:AVERage:COUNt:AUTO:MTIMe[?]
0.01 to 999.99
s
4.0
SENSe:AVERage:COUNt:AUTO:NSRatio[?]
0.0 to 1.0
dB
0.01
SENSe:AVERage:COUNt:AUTO:RESolution[?] 1 to 4
3
SENSe:AVERage:COUNt:AUTO:SLOT[?]
1 to <SENSe:POWer:TSLot:AVG
:COUNt>
1
SENSe:AVERage:COUNt:AUTO:TYPE[?]
RESolution | NSRatio
RESolution
SENSe:AVERage:RESet
SENSe:AVERage:STATe[?]
OFF | ON
ON
SENSe:AVERage:TCONtrol[?]
MOVing | REPeat
REPeat
SENSe:CORRection:DCYCle[?]
0.001 to 99.999
%
1.0
SENSe:CORRection:DCYCle:STATe[?]
OFF | ON
OFF
SENSe:CORRection:OFFSet[?]
-200.0 to 200.0
dB
0.0