Remote Control - Commands
R&S NRP-Z28/-Z98
1170.8966.12
6.8
E-4
SENSe:AVERage:COUNt[?] 1 to 65536
SENSe:AVERage:COUNt
sets the number of measured values that have to be averaged for forming
the measurement result in the modes
Continuous Average
,
Burst Average
and
Timeslot Average
. The
higher this averaging factor, the less the measured values fluctuate and the longer the measurement
time lasts. The parameter is rounded off to the nearest power-of-two number.
The query yields the averaging factor used in the modes
Continuous Average
,
Burst Average
and
Timeslot Average
.
The averaging function must be activated with
SENSe:AVERage:STATe ON
so that the
set averaging factor becomes effective.
Default setting:
4
SENSe:AVERage:COUNt:AUTO[?] OFF | ON | ONCE
SENSe:AVERage:COUNt:AUTO
activates (auto-averaging) or deactivates (fixed-filter mode) automatic
determination of the averaging factor in the modes
Continuous Average
,
Burst Average
and
Timeslot
Average
. If auto-averaging is activated, the averaging factor is continuously determined and set
depending on the level of power and other parameters.
SENSe:AVERage:COUNt:AUTO ON
activates
auto-averaging
and
SENSe:AVERage:COUNt
:AUTO OFF
deactivates it. On deactivation, the previous, automatically determined averaging factor is
used in the fixed-filter mode. The
SENSe:AVERage:COUNt:AUTO ONCE
command ensures that a new
averaging factor is determined by the filter automatic function under the current measurement
conditions and used in the fixed-filter mode.
The query yields
•
1
for
OFF,
•
2
for
ON
.
Default setting:
ON
SENSe:AVERage:COUNt:AUTO:MTIMe[?] 0.01 to 999.99
SENSe:AVERage:COUNt:AUTO:MTIMe
sets an upper limit for the settling time of the auto-averaging
filter in the
NSRatio
mode and thus limits the length of the filter.
The query returns the time that has been set.
Default setting:
4.0 [s]