Remote Control - Commands
R&S NRP-Z28/-Z98
1170.8966.12
6.18
E-4
Fig. 6-2
Correction of interactions between the power sensor and the DUT
If the gamma correction is made in connection with an s-parameter correction (setting
SENSe:CORRection:SPDevice:STATe ON
), the interaction of the DUT with the s-parameter device on
the one hand and the output of the sensor on the other hand (depending on the magnitude of the term
|
s
12
s
21
|
) is corrected. The interaction between the equivalent output reflection coefficient
eq
of the
sensor and the parameter
s
22
is always taken into account when the s-parameter correction is activated
– independent of the gamma correction in connection with the DUT.
Fig. 6-3
Correction of interactions between the power sensor, the DUT, and the s-parameter device
The query yields
•
1
for
OFF,
•
2
for
ON.
Default setting:
OFF
SENSe:SGAMma:MAGNitude[?] 0.0 to 1.0
SENSe:SGAMma:MAGNitude
defines the magnitude of the complex reflection coefficient of the DUT.
A value of
0.0
corresponds to an ideal match and a value of 1.0 to total reflection.
The query yields the set magnitude.
Default setting:
0.0
SENSe:SGAMma:PHASe[?] –360.0 to 360.0
SENSe:SGAMma:MAGNitude
defines the phase angle (in degrees) of the complex reflection coefficient
of the DUT.
The query yields the set phase angle.
Default setting:
0.0 [°]