R&S NRP-Z28/-Z98
Remote Control - Commands
1170.8966.12
6.23
E-4
SYSTem
With the aid of the SYSTem system, administrative device settings can be defined and queried. This
includes detailed information on the sensor and its initialization and the transfer of available commands
and their parameter limits.
Table 6-4
Commands of the
SYSTem
system
Command
Parameter
Unit
Remarks
SYSTem
:INFO? [Item]
Query only
:INITialize
No query
:MINPower?
W
Query only
:RUTime[?]
0.0 to 10.0
s
:SUTime[?]
0.0 to 10.0
s
:TRANsaction
:BEGIN:
No query
:END
No query
SYSTem:INFO? [Item]
SYSTem:INFO?
yields a string containing information that is more detailed than the identification string
delivered by the sensor as a response to
*IDN?
. If no
Item
is specified, the response string is a
sequence of entries in the form
Item:Information-String
separated by
CR
and
LF
(in C notation:
\r\n
).
With the
Item
optionally appended to the command, the entry for the required
Item
can be queried. The
response string is zero-terminated, i.e. its end identification is a zero byte (in C notation:
\0
).