R&S NRP-Z28/-Z98
Remote Control - Commands
1170.8966.12
6.19
E-4
SENSe:TIMing:EXCLude:STARt[?] 0.0 to 0.1
SENSe:TIMing:EXClude:STARt
defines an exclusion time at the start of a measurement window in the
Burst Average
and
Timeslot Average
modes. It is indicated by
T
START
in Figs. 6-4 and 6-5.
The query yields the exclusion time at the start of the measurement window.
Default setting:
0.0 [s]
SENSe:TIMing:EXCLude:STOP[?] 0.0 to –0.003
SENSe:TIMing:EXClude:STOP
defines an exclusion time at the end of a measurement window in the
Burst Average
and
Timeslot Average
modes. It is indicated by
T
STOP
in Figs. 6-4 and 6-5.
The query yields the exclusion time at the end of the measurement window.
Default setting:
0.0 [s]
Fig. 6-4
Effect of
SENSe:TIMing:EXCLude:STARt
and
:STOP
in the
Burst Average