Remote Control - Commands
R&S NRP-Z28/-Z98
1170.8966.12
6.16
E-4
Dropout time
Trigger
level
Power
Time
BURSt:DTOLerance
Pulse duration
BURSt:DTOLerance
Fig. 6-1
Effect of SENSe:POWer:BURSt:DTOLerance
SENSe:POWer:TSLot:AVG:COUNt[?] 1 to 128
For the Timeslot mode,
SENSe:POWer:TSLot:AVG:COUNt
sets the number of consecutive timeslots
that are to be processed after each trigger event.
The query yields the number of consecutive timeslots.
Default setting:
8
SENSe:POWer:TSLot:AVG:WIDTh[?] 10.0e-6 to 0.1
SENSe:POWer:TSLot:AVG:WIDTh
sets the length of a timeslot for the
Timeslot Average
mode (in s).
The query yields the length of a timeslot for the
Timeslot Average
mode.
Default setting:
0.001 [s]
SENSe:RANGe[?] 0 to 2
SENSe:RANGe
selects the measurement path of the sensor. The sensor has three separate
measurement paths. Path 1 is the most sensitive, path 2 medium and path 3 the least sensitive.
SENSe:RANGe 0
selects path 1,
SENSe:RANGe 1
path 2 and
SENSe:RANGe 2
path 3.
The dynamic ranges of these measurement paths depend on temperature and are model-specific.
Reference values are 40
S
W(–14 dBm) for the most sensitive measurement path, 4 mW (6 dBm) for the
medium one and 400 mW (26 dBm) for the least sensitive.
The query yields
•
0
for path 1,
•
1
for path 2,
•
2
for path 3.