Remote Control - Commands
R&S NRP-Z28/-Z98
1170.8966.12
6.32
E-4
TRIGger:SLOPe[?] POSitive | NEGative
TRIGger:SLOPe
defines the edge of the trigger event with internal and external triggering in the
Continuous Average, Timeslot Average,
and
Trace
modes. In this connection, positive means
increasing envelope power (with internal triggering) or increasing voltage (with external triggering). As in
the
Burst Average
mode, this command has no effect in conjunction with trigger sources
BUS
,
HOLD
and
IMMediate.
The query yields
•
1
for
POSitive
,
•
2
for
NEGative
.
Default setting:
POSitive
TRIGger:SOURce[?] HOLD | IMMediate | INTernal | BUS | EXTernal |
TRIGger:SOURce
sets the trigger source.
•
HOLD
:
Triggering only with command
TRIGger:IMMediate
.
•
IMMediate
: Automatic triggering without explicit event.
•
INTernal
:
Triggering by the measurement signal. Relevant parameters:
TRIGger:LEVel,
TRIGger:SLOPe, TRIGger:DELay, TRIGger:HYSTeresis,
TRIGger:HOLDoff
.
•
BUS
:
Triggering with command
*TRG
or
TRIGger:IMMediate
.
•
EXTernal
: Triggering via the hardware trigger bus, e.g. via the R&S NRP-Z3 USB adapter
or the R&S NRP base unit. Relevant parameters:
TRIGger:SLOPe, TRIGger:DELay, TRIGger:HOLDoff
.
The query yields
•
1
for
HOLD,
•
2
for
IMMediate,
•
4
for
INTernal,
•
8
for
BUS,
•
16
for
EXTernal.
Default setting:
IMMediate