Remote Control - Commands
R&S NRP-Z28/-Z98
1170.8966.12
6.26
E-4
SUTime
is usually set to a value that is slightly smaller than the response time of the control system.
Trigger events that have not yet occurred can then still be detected in time. At high trigger frequencies,
the result is that after the start of the measurement, only the first changeover to the
WAIT_FOR_TRIGGER
state and the subsequent changeover to the
MEASURING
state are reported.
The next message would not come until the transition to the IDLE state following the completion of the
measurement.
Default setting:
0.0001 [s]
SYSTem:TRANsaction:BEGin
SYSTEM:TRANsaction:BEGin
marks the beginning of a sequence of setting commands between which
the parameter limits must not be checked. This prevents the display of error messages when a setting
command causes a conflict that is resolved by a subsequent setting command. See
SYSTEM:TRANsaction:END
.
SYSTem:TRANsaction:END
SYSTEM:TRANsaction:END
marks the end of a sequence of setting commands between which the
parameter limits must not be checked. After this command, the parameter limits are checked.