R&S NRP-Z28/-Z98
Remote Control - Commands
1170.8966.12
6.11
E-4
SENSe:CORRection:DCYCle[?] 0.001 to 99.999
SENSe:CORRection:DCYCle
sets the duty cycle to a percent value for the correction of pulse-
modulated signals. With the correction activated, the sensor calculates the signal pulse power from this
value and the mean power. Since the duty cycle is only useful in the
Continuous Average
mode, it is
evaluated only there.
The query yields the current duty cycle in percent.
Default setting:
1.0 [%]
SENSe:CORRection:DCYCle:STATe[?] OFF | ON
SENSe:CORRection:DCYCle:STATe ON
activates the duty cycle correction and thus the pulse-power
measurement whereas
SENSe:CORRection:DCYCle:STATe OFF
deactivates it.
The query yields
•
1
for
OFF,
•
2
for
ON
.
Default setting:
OFF
SENSe:CORRection:OFFSet[?] –200.0 to 200.0
SENSe:CORRection:OFFSet
defines a fixed offset in dB, which is used to correct the measured value.
(When a log scale is used, the offset is added to the measured value; this is the reason why the
command has this name.)
An attenuator connected between the sensor and the DUT is taken into account with a negative offset.
In this case, the power detected by the sensor is reduced in line with the attenuation value, so that the
power is output at the input of the DUT. A positive offset could be used to correct the influence of a
gain.
The query yields the set offset in dB.
Default setting:
0.0 [dB]
SENSe:CORRection:OFFSet:STATe[?] OFF | ON
SENSe:CORRection:OFFSet:STATe ON
activates the offset correction and
SENSe:CORRection:OFFSet:STATe OFF
deactivates it.
The query yields
•
1
for
OFF,
•
2
for
ON.
Default setting:
OFF