166
CoaguChek
®
Pro II · Operator’s Manual
2
Clean the interior test strip guide.
Hold the meter upright with the test strip guide
facing down.
•
Clean the easily accessible area (shown in the
magnified illustration: white area below cotton
swab and its black surrounding frame) with a
water‑dampened cotton swab. Do not clean
beyond the white area inside the meter.
•
Ensure the swab is only damp, not wet. Wipe
away residual moisture and fluids.
•
Ensure the swab is only damp, not wet. Wipe
away residual moisture and fluids.
•
Carefully wipe the test strip guide dry using a
dry cotton swab.
Ensure that the test strip guide and cover are
completely dry before assembling them.
3
NOTICE!
Damage to the instrument
Do not insert any objects into the test strip guide.
Doing so could damage the electrical contacts
behind the test strip guide.
Close the cover and make sure it snaps into place.
8 Maintenance and Care
• 8
Содержание CoaguChek Pro II
Страница 1: ...CoaguChek Pro II Operator s Manual...
Страница 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Страница 29: ...29 CoaguChek Pro II Operator s Manual 1 3 Overview of the meter elements E F G H I J A B C D E 1 Introduction 1...
Страница 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 205: ......