171
CoaguChek
®
Pro II · Operator’s Manual
¾
The meter circuit board is defective. This can
happen when, for example, cleaning and
disinfecting agents enter the meter interior or
when there are dried residues of cleaning and
disinfecting agents in the meter interior.
¾
The test strip was wet or has already been used.
¾
The sample was applied too early, before warming
up of the strip was completed.
Additional information on
error E‑403
E‑403: Dosing Error
Error message
Touch
to exit this message. Perform the
suggested step(s) to solve the problem.
“Dosing” refers to the procedure of applying the blood
sample to the test strip. “Dosing Error” indicates that
an insufficient sample volume was detected. Please
check your procedure and repeat the test. Power the
meter off and remove the test strip. Repeat the
measurement using a new test strip and blood taken
from a new puncture site at the tip of another finger.
Make sure you apply a sufficient amount of blood to
the test strip.
If this error message appears again and again
although the blood sample was sufficient, the meter
might be contaminated with blood or residues of
cleaning and disinfecting agents.
9 Troubleshooting
• 9
Содержание CoaguChek Pro II
Страница 1: ...CoaguChek Pro II Operator s Manual...
Страница 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Страница 29: ...29 CoaguChek Pro II Operator s Manual 1 3 Overview of the meter elements E F G H I J A B C D E 1 Introduction 1...
Страница 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 205: ......