127
CoaguChek
®
Pro II · Operator’s Manual
5.3 Performing a liquid quality control test
r
To perform a liquid quality control test
1
Place the meter on a level, vibration‑free surface
or hold it in your hand so it is roughly horizontal.
2
Power the meter on by pressing
.
You can also power on the meter directly by
inserting a test strip or connecting the power
adapter.
3
Wait until the Main Menu is displayed, or log on as
described in
4
Check the battery level.
•
If the battery icon turns red (
one bar
remaining
), there may not be enough power
remaining for another test.
•
If there are
no bars
remaining in the battery
icon, you cannot perform any more tests.
Power the meter off by pressing
.
In both cases, restore power by recharging the
battery pack, or using the power adapter.
5
Check that the date and time are correct. Correct
any wrong entries as described in Chapter 3,
Meter Setup
.
5 Control and Proficiency Testing
• 5
Содержание CoaguChek Pro II
Страница 1: ...CoaguChek Pro II Operator s Manual...
Страница 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Страница 29: ...29 CoaguChek Pro II Operator s Manual 1 3 Overview of the meter elements E F G H I J A B C D E 1 Introduction 1...
Страница 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 205: ......