124
CoaguChek
®
Pro II · Operator’s Manual
5.2 Preparing to perform a liquid quality control test
Prepare for a liquid quality control test in the same
way you would prepare to perform a test with a
capillary blood sample. The only difference is the use
of control solution instead of blood.
r
To prepare for a liquid quality control test
1
Have the test strip container at hand.
2
If you are using the test strip lot for the first time,
make sure that the code chip that came with these
test strips is at hand.
3
Make sure the bottle of freeze‑dried (lyophilized)
control plasma and the dropper for making the
control solution are at hand. Always refer to
package inserts for information on the specified
operating conditions.
4
Make sure that the quality control code chip that
came with the control solution is at hand.
5
Open the lid of the bottle and remove the rubber
cap.
5 Control and Proficiency Testing
• 5
Содержание CoaguChek Pro II
Страница 1: ...CoaguChek Pro II Operator s Manual...
Страница 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Страница 29: ...29 CoaguChek Pro II Operator s Manual 1 3 Overview of the meter elements E F G H I J A B C D E 1 Introduction 1...
Страница 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Страница 205: ......