Страница 2: ...us Blood Glucose Monitoring System is comprised of the VivaChek Ino Plus Blood Glucose Meter and the VivaChek Ino Blood Glucose Test Strips The VivaChek Ino Plus Blood Glucose Monitoring System Meter...
Страница 3: ...esults wirelessly to the app 7 3 Taking a Test 8 Preparing the Test Strip 8 Getting a Blood Drop and Testing 9 Discard the Used Test Strip 10 Testing with Control Solution 11 Using the Meter Memory 14...
Страница 4: ...1 CHARPTER 1 UNDERSTANDING YOUR TESTING TOOLS Your Meter System Overview The VivaChek Ino Plus Blood Glucose Meter and Strip...
Страница 5: ...at the above operation All of the segments should be clear and exactly like the picture below If not contact VivaChek Customer Support at 800 951 8430 7 days a week 9 a m 5 p m Eastern Time for help P...
Страница 6: ...h beep sound function the meter will beep when turn on the meter the test strip is inserted and ready to apply blood or control solution sufficient blood or control solution is pulled into the test st...
Страница 7: ...est strips and Safety lancets are for single use only Do not drop blood directly on the flat surface of the test strip Check the expiration dates and discard dates on your test strips vial label or on...
Страница 8: ...m should not be used following xylose absorption procedures Not for neonatal use Not for screening or diagnosis of diabetes mellitus Note This device complies with part 15 of the FCC Rules Operation i...
Страница 9: ...e Before you first time use your meter system for testing please adjust the meter settings to set the date and time ensuring that results stored in the memory are shown with the correct date and time...
Страница 10: ...OTE The button will continue to blink with blue light after successful pairing Initial syncing of the meter and app You must initially sync the meter and app before you start testing your blood glucos...
Страница 11: ...on after a beep 4 A symbol with a test strip with a flashing blood will appear letting you know the meter is ready to test Note Check the expiration and discard dates on the test strip vial All expira...
Страница 12: ...rps container Please see the Caution Statement 4 Gently massage from the surrounding area toward the puncture site to collect the required blood volume For the greatest reduction in pain lance on the...
Страница 13: ...after a beep The test result will automatically be stored in the meter memory Please do not touch the test strip during the countdown as this may result in an error Discard the Used Test Strip You ca...
Страница 14: ...and the discard date of the test strip Make sure that the test strip vial has not been opened for more than 4 months Confirm the temperature in which you are testing is between 41 113 F Make sure that...
Страница 15: ...that is past the expiration date or discard date the control solution will expire 4 months after the bottle is opened for the first time Shake the bottle well before use Close the bottle tightly afte...
Страница 16: ...dd a second drop Discard the test strip and start over with a new test strip 6 Hold it in the drop until the meter beeps and then you see the meter count down on the screen and followed with your cont...
Страница 17: ...king properly DO NOT use the system to test blood Contact VivaChek Customer Support at 800 951 8430 7 days a week 9 a m 5 p m Eastern Time for help Please contact your health care professional if you...
Страница 18: ...r or poison control centre Discard batteries according to your local environmental regulations Caring for Your Glucose Monitoring System Store meter in the carrying case provided whenever possible Was...
Страница 19: ...th motion for a total of 4 motions And then please expose the meter to air dry for at least 2 minute as specified on the disinfectant wipe manufacturer s label Disinfection is performed using one wipe...
Страница 20: ...t with a new test strip Wait until you see the flashing blood drop on the display before testing E2 The meter is sensing a used or contaminated test strip Discard the test strip and repeat the test wi...
Страница 21: ...epeat the test using a new test strip If your result still flashes HI contact your health care professional as soon as possible Caution Glucose levels above 250 mg dL 13 9 mmol L may indicate a potent...
Страница 22: ...symptoms test your blood glucose If your test result is under 50 mg dL or above 250 mg dL contact your health care professional immediately Low blood glucose Hypoglycemia shakiness sweating fast hear...
Страница 23: ...teries Glucose units of measure The meter is pre set to milligrams per deciliter mg dL Memory Up to 1000 records with date and time Automatic shutoff 2 minutes after last action Dimensions 90 4 mm x 5...