60 • VOC Pro Transmitter
2. Press the ADD button to begin the zero process and advance to the clean air confirmation
screen.
3. If the sensor is in clean air, press the ADD button to select “Yes” and continue to Step 5.
4. If the sensor is not in clear air:
a. Unscrew and remove the rain guard from the assembly.
b. Install the calibration cup to the VOC Pro’s sensor housing.
c. Screw the regulator into the zero air calibration cylinder.
d. Use the sample tubing to connect the regulator to the calibration cup.
e. Turn the regulator knob counterclockwise to open the regulator.
f. Allow the gas to flow for two minutes.
g. Press the ADD button to select “Yes” and continue to Step 5.
5. The unit will automatically begin the 6-second zero process. During the zero, the display
will show a countdown of the time remaining until the process is complete.
NOTE:
The zero process cannot be stopped without disconnecting the power from the unit.
6. When the zero process is complete, press the MENU button to advance to the calibration
screen.
7. If a zero air calibration cylinder was used, turn the regulator knob clockwise to close the
regulator.