VOC Pro Transmitter • 47
6. Hold the PID sensor steady on the working surface with one hand and using the other
hand, locate the tabs on the electrode stack removal tool and insert them into the slots on
the side of the PID sensor near the face.
Figure 9: Using Removal Tool
7. Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack and
lamp are released.
8. Carefully lift the PID sensor body away from the electrode stack and lamp. Take care not
to touch the lamp window, the flat end of the lamp, with your fingers. If the lamp remains
lodged in the sensor body, carefully remove it with tweezers.
Figure 10: Lamp Window Location
9. If the spring in the lamp cavity comes out, place it back into the lamp cavity.
10. Hold the lamp in one hand being careful not to touch the lamp window with your fingers.
11. With the other hand collect a small amount of powder (aluminum oxide for 10.0 eV and
10.6 eV sensors; anhydrous methanol for 11.7 eV sensors) on a cotton swab.
Lamp Window