VOC Pro Transmitter • 51
5. Hold the PID sensor steady on the working surface with one hand and using the other
hand, locate the tabs on the electrode stack removal tool and insert them into the slots on
the side of the PID sensor near the face.
Figure 13: Using Removal Tool
6. Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack and
lamp are released.
7. Carefully lift the PID sensor body away from the electrode stack and lamp. If the lamp
remains lodged in the sensor body, carefully remove it with tweezers.
8. If the spring in the lamp cavity comes out, place it back into the lamp cavity.
9. Discard the old PID lamp.
NOTE:
At this point it is recommended that the finger cots be used on the fingers handling the
lamp. Finger cots are included with the lamp cleaning kit.
10. Hold the electrode stack between the thumb and forefinger of one hand and place the
window end of the new lamp inside the O-ring seal in the electrode stack with the other
hand as shown below. Take care not to touch the lamp window.
Figure 14: Reinstalling the Electrode Lamp
11. Twisting the lamp slightly during insertion will help to ensure the lamp window is snug
against the stack's front electrode. The lamp should be supported by the O-ring.
12. Continuing to hold the electrode stack between your forefinger and thumb, carefully insert
the lamp into the lamp cavity in the sensor ensuring that the lamp remains in position.