18 • VOC Pro Transmitter
3. Connect the RS-485 B (BROWN) wire to the
B
terminal.
4. Connect the ground (WHITE) wire to the
GND
terminal.
5. Connect the RS-485 A (YELLOW) wire to the
A
terminal.
6. Feed the RS-485 cable through the power hub and into the controller and wire them to the
correct terminals as shown below.
Figure 6: Modbus Wiring
NOTE:
If an RKI controller is not used, the VOC Pro can be connected to a Programmable
Logic Controller (PLC) for RS-485 Modbus data communications. For integration
and setup, refer to the Modbus Register Map found in Appendix C of this manual.