46 • VOC Pro Transmitter
When using the anhydrous methanol for the 11.7 eV sensor, try to minimize the vial’s
exposure to air by leaving the cap on as much as possible. Discard the anhydrous methanol 3
months after opening or after leaving the cap off for an extended period of time.
CAUTION:
Using the anhydrous methanol more than 3 months after opening it or after the
cap has been off for an extended period of time may damage the 11.7 eV lamp.
Lamp Cleaning Procedure
1. Press and hold the SUB button for approximately 6 seconds, until “OFF” shows on the
display screen.
2. Unscrew and remove the sensor housing cap from the sensor housing base. Set aside.
3. Gently unplug the sensor element from the sensor housing board.
4. Place the PID sensor face down on a flat clean working surface.
5. Do not touch the lamp window with your fingers as this may contaminate the window
with finger oil. At this point it is recommended that the finger cots be used on the fingers
handling the lamp. Finger cots are included with the lamp cleaning kit.